LeapSecond.com :: Museum of Time & Frequency |
|
Sigma Tau MHM-2010 Hydrogen Maser |
|
Topics: H-maser, Active Hydrogen Maser, Passive Hydrogen Maser, Allan Deviation, Frequency Stability, 3-Cornered Hat
Here's the first results of comparisons among three masers.
Comparator: TSC 5110A S/N 036 upper lab
(ch A) ST = Sigma Tau Active Hydrogen Maser MHM-2010 S/N 208
(ch B) PHM = KVARZ CH1-76 Passive Hydrogen Maser
TSC5110A plot: ST to PHM after 60k seconds (17 hours) |
Comparator: TSC 5110A S/N 006 lower lab
(ch A) AHM = KVARZ CH1-75 Active Hydrogen Maser
(ch B) ST = Sigma Tau Active Hydrogen Maser MHM-2010 S/N 208
TSC5110A plot: AHM to ST after 60k seconds (17 hours) |
At the same scale (5E-13/div) note AHM short-term noise is a fraction of PHM: